    Six major medical device regulators compared on nine topics. Each cell is a short summary; follow the linked terms for the full definition and primary sources. Always confirm details against the regulator's current text.

    Take FDA, EU and TGA quiz
    Topic
    United States
    FDA (CDRH)
    European Union
    United Kingdom
    MHRA
    Canada
    Japan
    Australia
    Core law
    Federal Food, Drug, and Cosmetic Act; device rules in 21 CFR Parts 800 to 1299.
    Regulation (EU) 2017/745 (MDR), applying since May 26, 2021.
    Medical Devices Regulations 2002 (SI 2002/618), as amended for Great Britain.
    Medical Devices Regulations (SOR/98-282) under the Food and Drugs Act.
    Act on Pharmaceuticals and Medical Devices (PMD Act).
    Therapeutic Goods Act 1989 and Therapeutic Goods (Medical Devices) Regulations 2002.
    Risk classes
    Class I, II, III.
    Class I (including Is, Im, Ir), IIa, IIb, III under Annex VIII rules.
    Class I, IIa, IIb, III.
    Class I, II, III, IV.
    Class I, II, III, IV.
    Class I (including Is, Im), IIa, IIb and III. Active implantable medical devices are Class III.
    Route to market
    Exemption, 510(k), De Novo or PMA depending on class and predicate.
    Conformity assessment, with a notified body for most classes above I, then CE marking.
    Great Britain: UKCA, with a UK approved body where required; non-sterile, non-measuring Class I devices can self-declare. Published CE transitions: eligible MDD/AIMDD devices until certificate expiry or June 30, 2028, whichever is sooner; eligible IVDD devices until certificate expiry or June 30, 2030, whichever is sooner; MDR/IVDR devices until June 30, 2030. Certificate validity and device-specific conditions apply. On October 21, 2026, the government announced indefinite CE recognition; implementation details should be checked, not treated as already replacing the published transitions. Northern Ireland: CE under EU rules (CE + UKNI where a UK body is used); UKCA alone is not accepted.
    Medical Device Licence (MDL) for Classes II, III, IV. Class I does not require an MDL. An MDEL licenses the establishment, not the product: generally required for importers/distributors of all classes and Class I manufacturers, subject to exemptions (including Class I manufacturers selling solely through an MDEL holder).
    Class I: notification to PMDA. Class II and III: registered-body certification where certification standards exist; otherwise MHLW approval after PMDA review. Class IV: MHLW approval after PMDA review. Foreign manufacturers proceed through a Japanese marketing authorisation holder or appointed Japanese manufacturer.
    Inclusion in the ARTG by an Australian sponsor, often using comparable overseas approvals.
    Quality system
    QMSR (21 CFR Part 820), incorporating ISO 13485:2016, effective February 2, 2026.
    QMS required by MDR Article 10, usually shown through EN ISO 13485.
    ISO 13485 based QMS assessed by the approved body.
    ISO 13485 certificate issued under MDSAP for Class II to IV.
    MHLW QMS Ministerial Ordinance, aligned with ISO 13485; MDSAP reports accepted.
    ISO 13485 based QMS; MDSAP audit reports used as evidence.
    Technical file and clinical evidence
    Submission content set by pathway; significant risk studies need an IDE.
    Technical documentation (Annexes II and III) and clinical evaluation with PMCF.
    Technical documentation and clinical evaluation reviewed by the approved body.
    Licence application with safety and effectiveness evidence scaled to class.
    Application dossier (STED format) with clinical data where required.
    Essential Principles checklist and clinical evidence held by the sponsor.
    Post-market and vigilance
    Medical Device Reporting (21 CFR Part 803), corrections and removals, recalls.
    PMS system, vigilance (Articles 87 to 92), PSUR, FSCA reporting.
    Great Britain: strengthened Part 4A PMS rules (SI 2024/1368) apply to individual devices placed on the market or put into service from June 16, 2025, including CE- and UKCA-marked devices, subject to scope exclusions. Earlier individual devices remain under prior PMS rules, even where later units of the same model fall under the new rules. Northern Ireland follows EU MDR/IVDR PMS rules.
    Mandatory problem reporting and recall reporting under the Regulations.
    Adverse event and malfunction reporting to PMDA under the PMD Act.
    Adverse event reporting by the sponsor to the TGA.
    Device identification
    UDI on labels and data in GUDID (21 CFR Part 830).
    UDI system with EUDAMED registration. Actor, UDI/Device, Notified Bodies & Certificates, Market Surveillance modules became mandatory May 28, 2026, subject to transitional provisions. Market Surveillance is for competent authorities and the Commission. Other modules are not yet mandatory.
    Great Britain: MHRA device registration; future UDI requirements are part of regulatory reform, with no mandatory commencement date verified in this audit. Northern Ireland follows applicable EU MDR/IVDR UDI requirements.
    Official regulator document

    Source checked:

    Health Canada consulted on a proposed UDI system from June 28 to September 30, 2021. No enacted Canadian UDI mandate or compliance date was verified in this audit.
    Official regulator document

    Source checked:

    Mandatory identification-code labelling for in-scope medical devices and IVDs began December 1, 2022, using GS1-based codes, subject to exceptions and transitional provisions.
    UDI labelling and AusUDID data requirements began July 1, 2026 for Class IIb and III devices, subject to scope exclusions and transitional arrangements. Other in-scope classes are phased in later.
    Cybersecurity
    Section 524B of the FD&C Act for cyber devices, including an SBOM, since March 29, 2023.
    Annex I sections 17.2 and 17.4 (IT security and protection against unauthorised access), interpreted through MDCG 2019-16 rev.1.
    Assessed against essential requirements; follows international guidance.
    Health Canada premarket cybersecurity guidance.
    Cybersecurity expectations aligned with IMDRF guidance.
    Essential Principles in Schedule 1 of the Medical Devices Regulations 2002, interpreted through TGA cyber security guidance for industry.
    Local representative
    US agent required for foreign establishments.
    EU authorised representative for non-EU makers, plus a PRRC.
    UK Responsible Person for makers outside the UK.
    No formal representative; importers and distributors need establishment licences.
    Marketing authorisation holder in Japan; foreign makers designate one.
    Australian sponsor holds the ARTG entry.