US FDA vs EU MDR, side by side
The US and the EU reach the same goal, safe and effective devices, through different machinery. In the US a single government agency (FDA) reviews and clears or approves devices. In the EU the manufacturer self-declares conformity, and for most devices a private, government-designated notified body audits that claim before the CE mark can be applied.
By Christian Espinosa, Founder, Blue Goat Cyber
Who decides and how
| Topic | United States (FDA) | European Union (MDR) | What differs |
|---|---|---|---|
| Legal basis | Federal Food, Drug, and Cosmetic Act and 21 CFR Parts 800 to 89821 CFR Subchapter H | Regulation (EU) 2017/745, directly applicable in every Member StateMDR | Both are binding law. The MDR replaced the earlier directives and applies since 26 May 2021. |
| Who grants market access | FDA (CDRH) clears, grants or approves the deviceFD&C Act sections 510(k), 513(f)(2), 515 | Manufacturer declares conformity; a notified body certifies most devicesMDR Articles 52 and 53 | In the US the regulator reviews the file. In the EU a designated private body audits it, and national competent authorities supervise the market. |
| Evidence of market access | 510(k) clearance letter, De Novo grant or PMA approvalFDA databases | CE marking backed by an EU declaration of conformityMDR Articles 19 and 20 | A CE mark is valid across the EU and EEA; an FDA decision covers the US only. |
Classification and pathways
| Topic | United States (FDA) | European Union (MDR) | What differs |
|---|---|---|---|
| Risk classes | Class I, II, III, set by product classification regulations21 CFR Parts 862 to 892 | Class I, IIa, IIb, III, set by 22 rulesMDR Annex VIII | US classes come from a lookup by device type; EU classes come from applying rules (invasiveness, duration, active or not). The same device can land in different classes. |
| Moderate-risk route | 510(k), showing substantial equivalence to a predicate21 CFR Part 807 Subpart E | Conformity assessment with notified body (Class IIa and IIb)MDR Article 52, Annexes IX to XI | The EU has no predicate concept. Equivalence can support clinical evaluation, but the device must meet the GSPRs on its own evidence. |
| Novel low or moderate risk | De Novo classification requestFD&C Act section 513(f)(2) | No separate pathway; classify by Annex VIII rulesMDR Annex VIII | De Novo creates a new classification regulation that later devices can use as a predicate. |
| Highest risk | Premarket approval (PMA)21 CFR Part 814 | Class III notified body assessment, with expert panel consultation for some implantablesMDR Article 54 | Both require clinical data. PMA approval is device-specific and changes need PMA supplements. |
| Early regulator feedback | Q-Submission (Pre-Sub)FDA Q-Submission program | Structured dialogue with the notified body; expert panel adviceMDR Article 61(2), Article 106 | Notified bodies may not provide consultancy, so EU early dialogue is narrower than an FDA Pre-Sub. |
Quality and technical documentation
| Topic | United States (FDA) | European Union (MDR) | What differs |
|---|---|---|---|
| Quality management system | QMSR, which incorporates ISO 13485:2016 by reference (effective 2 February 2026)21 CFR Part 820 | QMS required by MDR, usually demonstrated through ISO 13485MDR Article 10(9) | With QMSR, the US and EU now share the same core QMS standard, though each adds its own requirements. |
| Design and risk | Design controls and risk management21 CFR Part 820, ISO 13485 clause 7.3 | GSPRs plus risk management, usually via ISO 14971MDR Annex I | ISO 14971 is recognized by FDA and harmonised in the EU, so one risk file can serve both. |
| Technical file | Submission content (510(k), De Novo, PMA) plus the design history21 CFR 807.87, 814.20 | Technical documentation per Annexes II and IIIMDR Annexes II and III | The EU file is a standing, maintained document; the US file is mainly what is submitted for each decision. |
| Responsible person | No equivalent single named role- | Person responsible for regulatory compliance (PRRC)MDR Article 15 | Non-EU manufacturers also need an EU authorised representative (MDR Article 11). |
Clinical evidence
| Topic | United States (FDA) | European Union (MDR) | What differs |
|---|---|---|---|
| Clinical studies | Investigational Device Exemption (IDE) for significant-risk studies21 CFR Part 812 | Clinical investigations authorised by Member StatesMDR Articles 62 to 82, ISO 14155 | Both follow ISO 14155 and good clinical practice principles. |
| Ongoing clinical evaluation | No standing requirement for most cleared devices- | Clinical evaluation updated with post-market clinical follow-upMDR Article 61, Annex XIV | The EU expects clinical evidence to be updated for the life of the device. |
| Public summary | 510(k) summary, De Novo decision summary, PMA SSEDFDA databases | Summary of safety and clinical performance (Class III and implantables)MDR Article 32 | The SSCP is published through EUDAMED. |
Identification and post-market
| Topic | United States (FDA) | European Union (MDR) | What differs |
|---|---|---|---|
| Unique device identification | UDI with data in GUDID21 CFR Part 830 | UDI with data in EUDAMED, plus Basic UDI-DIMDR Article 27 | Both use the same issuing agencies (for example GS1), but data elements differ. |
| Adverse event reporting | Medical Device Reporting (MDR) to FDA21 CFR Part 803 | Vigilance reporting of serious incidentsMDR Articles 87 to 92 | Watch the acronym clash: in US usage MDR means Medical Device Reporting. |
| Field actions | Recalls and corrections and removals21 CFR Parts 7 and 806 | Field safety corrective actions (FSCA)MDR Article 89 | Both require notifying the regulator and affected users. |
| Periodic safety reporting | PMA annual reports; no general periodic report for 510(k) devices21 CFR 814.84 | PSUR (Class IIa and above) or PMS report (Class I)MDR Articles 85 and 86 | The EU requires periodic summaries for most device classes. |
Primary sources
- 21 CFR Part 820 (QMSR)
- 21 CFR Part 807 (510(k))
- 21 CFR Part 814 (PMA)
- 21 CFR Part 803 (Medical Device Reporting)
- Regulation (EU) 2017/745 (MDR)
Educational summary, not legal advice. Always check the current text of the law and guidance.