ISO 17665
Standard for moist heat (steam) sterilization of medical devices.
Reviewed by Christian Espinosa, Founder, Blue Goat CyberLast reviewed May 5, 2026
Definition
ISO 17665 specifies the requirements for development, validation, and routine control of a moist-heat sterilization process. The 2024 revision consolidated industrial and healthcare-facility requirements. What the regulation says
ISO 17665 is recognized by regulatory bodies worldwide as a key standard for moist-heat sterilization, particularly for medical devices, as noted by the FDA in its guidance on reprocessing medical devices. Compliance with this standard helps manufacturers and healthcare facilities demonstrate that their sterilization processes are effective and consistently achieve sterility, which is essential for meeting the safety requirements of regulations like the EU MDR, Annex I, Section 17.2.
What this means in practice
Steam sterilization is the workhorse of reusable surgical instruments and is referenced in reprocessing instructions for many devices.Examples
- A MedTech manufacturer developing a new reusable surgical instrument must validate its moist-heat sterilization instructions according to ISO 17665 before placing the device on the market.
- A hospital central sterile supply department (CSSD) uses ISO 17665 as a framework to ensure that its steam sterilizers are correctly operated and routinely monitored for effective sterilization of surgical trays.
- An auditor reviewing a medical device manufacturer's quality management system would expect to see documented evidence of sterilization process validation in line with ISO 17665, including reports on physical, chemical, and biological indicators.
Common pitfalls
- •A common pitfall is failing to re-validate sterilization processes after changes to device design, packaging materials, or sterilization equipment.
- •Another mistake is neglecting to establish clear process parameters and acceptance criteria for routine monitoring, leading to inconsistencies in sterilization outcomes.
- •Many organizations incorrectly assume that simply purchasing ISO 17665 compliant equipment guarantees a validated process, overlooking the need for site-specific validation studies.
- •Inadequate documentation of validation activities and routine monitoring records often results in non-compliance during regulatory audits.
- •Failing to define and implement appropriate corrective actions when sterilization process deviations occur is a significant oversight.
Frequently asked questions
The primary purpose is to specify requirements for the development, validation, and routine control of sterilization processes for medical devices using moist heat, ensuring the sterility and safety of reprocessed items.
Related terms
Grouped by themeEditor's picks
· Hand-selected related conceptsMore in Standards
· Same categoryStandards
ASTM F2503
Standard practice for marking medical devices and other items for safety in the magnetic resonance environment.
Standards
Essential Performance
Performance of a clinical function whose loss or degradation would result in unacceptable risk.
Standards
ICH E6(R3) Good Clinical Practice(E6(R3))
Revision 3 of the ICH Good Clinical Practice guideline, restructured around principles, modernized for risk-based and decentralized trials, finalized in 2023.
Standards
IEC 60601-1
General requirements for basic safety and essential performance of medical electrical equipment.
Primary references
3 sourcesLink health: 3 verified· last checked 2026-06-20
ISO·2FDA·1
- 1
ISO 17665VerifiedISOiso.org
- 2
FDA Recognized Consensus StandardsVerifiedFDAaccessdata.fda.gov
- 3
ISO Standards Catalogue - HealthVerifiedISOiso.org
Inline markers like [1] jump to the matching reference above.