    ASTM F2503

    Standard practice for marking medical devices and other items for safety in the magnetic resonance environment.

    Reviewed by Christian Espinosa, Founder, Blue Goat CyberLast reviewed September 19, 2026

    Definition

    ASTM F2503 defines the MR Safe (green square), MR Conditional (yellow triangle), and MR Unsafe (red circle) markings used on devices and accessories in MR-equipped facilities. MR Conditional devices must have labeling that describes the conditions for safe scanning, such as static magnetic field strength, spatial field gradient, RF exposure (specific absorption rate) and expected heating and image artifacts. Test methods are set out in related ASTM standards, such as F2052 (magnetically induced force), F2213 (torque), F2182 (RF heating) and F2119 (artifacts). FDA's guidance on MR safety labeling references ASTM F2503.
    What the regulation says
    ASTM F2503, Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance (MR) Environment, provides a standardized system for labeling medical devices and other items to indicate their compatibility with the MR environment. This standard is recognized by regulatory bodies such as the FDA in their guidance documents concerning MR compatibility of medical devices.

    What this means in practice

    Required labeling element for any device that may enter or be used near MR scanners, including implants, monitoring equipment, and accessories.

    Examples

    • A pacemaker is labeled 'MR Conditional' with specific conditions, including a maximum static magnetic field strength of 1.5 Tesla and a maximum spatial gradient field of 500 Gauss/cm.
    • A patient undergoing an MRI has a metallic joint implant, which is correctly identified as 'MR Conditional' based on its labeling, and the imaging protocol is adjusted accordingly.
    • A hospital purchases new infusion pumps that are clearly marked 'MR Safe,' allowing them to be used directly within the MRI scanner room without concern for magnetic interference.
    Common pitfalls
    • •Misinterpreting 'MR Conditional' to mean a device is safe under all MR conditions, without considering the specific parameters and warnings outlined in the device's labeling.
    • •Failing to properly train personnel on the meaning of MR safety labels and the associated risks.
    • •Applying incorrect or unverified MR safety labels to devices, leading to potential harm to patients or damage to equipment.
    • •Not accounting for changes in a device's MR compatibility after repair or modification.

    Frequently asked questions

    The standard defines three primary classifications: MR Safe (green square), MR Conditional (yellow triangle), and MR Unsafe (red circle). These icons visually communicate the device's compatibility with magnetic resonance environments.
    Grouped by theme
    Cited by

    Where this term appears across MedTech Terms.

    Latest in MedTech

    Sources

    3 sources

    Sources support the definition or relevant regulatory context, not necessarily every claim on this page. Each is graded against our source-tier hierarchy so you can see what rests on binding law versus commentary.

    Tier 1Binding law and standards· 1Tier 2Regulator guidance and consensus· 2
    Link health: 3 verified· last checked 2026-06-20
    ASTM·1AAMI·1FDA·1
    1. 1
      ASTM F2503
      Tier 1 Verified
      ASTMastm.org
    2. 2
      AAMI Standards
      Tier 2 Verified
      AAMIaami.org
    3. 3
      FDA Recognized Consensus Standards
      Tier 2 Verified
      FDAaccessdata.fda.gov

    Inline markers like [1] jump to the matching reference above.