ASTM F2503
Standard practice for marking medical devices and other items for safety in the magnetic resonance environment.
Definition
ASTM F2503 defines the MR Safe (green square), MR Conditional (yellow triangle), and MR Unsafe (red circle) markings used on devices and accessories in MR-equipped facilities. MR Conditional devices must have labeling that describes the conditions for safe scanning, such as static magnetic field strength, spatial field gradient, RF exposure (specific absorption rate) and expected heating and image artifacts. Test methods are set out in related ASTM standards, such as F2052 (magnetically induced force), F2213 (torque), F2182 (RF heating) and F2119 (artifacts). FDA's guidance on MR safety labeling references ASTM F2503.What this means in practice
Required labeling element for any device that may enter or be used near MR scanners, including implants, monitoring equipment, and accessories.Examples
- A pacemaker is labeled 'MR Conditional' with specific conditions, including a maximum static magnetic field strength of 1.5 Tesla and a maximum spatial gradient field of 500 Gauss/cm.
- A patient undergoing an MRI has a metallic joint implant, which is correctly identified as 'MR Conditional' based on its labeling, and the imaging protocol is adjusted accordingly.
- A hospital purchases new infusion pumps that are clearly marked 'MR Safe,' allowing them to be used directly within the MRI scanner room without concern for magnetic interference.
- •Misinterpreting 'MR Conditional' to mean a device is safe under all MR conditions, without considering the specific parameters and warnings outlined in the device's labeling.
- •Failing to properly train personnel on the meaning of MR safety labels and the associated risks.
- •Applying incorrect or unverified MR safety labels to devices, leading to potential harm to patients or damage to equipment.
- •Not accounting for changes in a device's MR compatibility after repair or modification.
Frequently asked questions
Related terms
Grouped by themeEditor's picks
· Hand-selected related conceptsMore in Standards
· Same categoryPerformance of a clinical function whose loss or degradation would result in unacceptable risk.
European standards cited by reference in the Official Journal of the EU that give manufacturers a presumption of conformity with corresponding legal requirements.
ICH Good Clinical Practice revision with Principles and Annex 1 adopted in January 2025 and Annex 2 in June 2026.
General requirements for basic safety and essential performance of medical electrical equipment.
Where this term appears across MedTech Terms.
Latest in MedTech
Sources
3 sourcesSources support the definition or relevant regulatory context, not necessarily every claim on this page. Each is graded against our source-tier hierarchy so you can see what rests on binding law versus commentary.
- 1ASTM F2503Tier 1 VerifiedASTMastm.org
- 2AAMI StandardsTier 2 VerifiedAAMIaami.org
- 3FDA Recognized Consensus StandardsTier 2 VerifiedFDAaccessdata.fda.gov
Inline markers like [1] jump to the matching reference above.