    Post-Market

    Post-Market Surveillance(PMS)

    In one line
    Activities to monitor device safety and performance after market release.
    Definition
    Post-Market Surveillance (PMS) comprises the systematic processes a manufacturer establishes to actively collect and review experience gained from devices placed on the market, in order to identify any need to apply corrective or preventive actions. Under the EU MDR (Articles 83 to 86) and IVDR (Articles 78 to 81), every manufacturer must run a PMS system proportionate to risk class, described in a PMS plan, with outputs in a PMS report or PSUR. In the US, PMS is spread across complaint handling, medical device reporting, corrections and removals and, for some devices, FDA-ordered section 522 studies. PMS data must feed back into risk management, clinical evaluation and CAPA.
    Why it matters
    PMS is required under EU MDR Articles 83-86 and IVDR Articles 78-81, and is embedded in FDA's Quality System Regulation (soon QMSR) through complaint handling (820.198), CAPA (820.100), and MDR reporting (Part 803). Under EU MDR, manufacturers must maintain a PMS plan proportionate to device risk (Article 84), covering active and reactive data collection: vigilance and serious incidents, trend reports (Article 88), user feedback, scientific literature, similar-device data, PMCF for Class IIa and above, and PMS reports (Class I) or PSURs (Class IIa+). Outputs feed the risk management file, technical documentation, clinical evaluation, and, where needed, corrective actions and field safety corrective actions (FSCA).
    Common pitfalls
    • •Confusing PMS (proactive, plan-driven) with vigilance (reactive, incident-driven). MDR Article 83 explicitly distinguishes them; you need both.
    • •Under-scoping PMS to complaints only. MDR expects trend analysis, literature review, similar-device data, and PMCF results as well.
    • •Missing the trend reporting obligation. MDR Article 88 requires reporting statistically significant increases in non-serious incidents or expected side effects.
    • •Treating PMS as separate from the RMF. PMS outputs must be linked to risk analysis updates and, where relevant, to CAPA.
    Open full page
    Select term B to compare.

    Why compare MedTech terms side by side?

    MedTech terminology is full of pairs that look interchangeable but carry very different regulatory, clinical, and commercial consequences. Picking the wrong framework, pathway, or standard early in a project can add months to a submission, invalidate clinical evidence, or trigger an audit finding. Side-by-side comparison is the fastest way to surface those differences before they become costly mistakes.

    Each comparison on this page pulls from the same vendor-neutral, sourced definitions used throughout the MedTech Terms glossary. You see the one-line summary, the formal definition, why it matters in practice, common pitfalls, and the primary sources (FDA guidance, EU MDR/IVDR articles, ISO/IEC standards, MDCG documents, IMDRF principles) that back each entry. That makes the comparison defensible in regulatory strategy memos, design reviews, and submission narratives.

    Common comparison patterns

    How to use this tool

    Pick term A and term B from the dropdowns, or click a preset above. The URL updates with both slugs so you can bookmark or share the exact comparison with a colleague, a notified body reviewer, or your regulatory consultant. Click Open full page on either side for the complete entry, including FAQs, related terms, and the full citation list. If you are not sure which term to start with, browse the Categories view or the A-Z index.