    Quality & Risk

    Verification & Validation(V&V)

    In one line
    Confirming design outputs meet inputs (V) and that devices meet user needs (V).
    Definition
    Verification confirms by examination that specified requirements have been fulfilled (did we build the device right?). Validation confirms that requirements for a specific intended use can be consistently fulfilled (did we build the right device?). Verification is usually done through testing, inspection and analysis against design inputs, such as bench tests, software unit and integration tests, and electrical safety testing. Validation checks the finished device, or an equivalent, under actual or simulated use conditions with intended users, and includes software validation, human factors validation and, where needed, clinical evaluation. Both must be planned, use predefined acceptance criteria and be documented.
    Why it matters
    V&V plans, protocols, and reports are central artifacts in design controls and submissions. Software V&V follows IEC 62304.
    Common pitfalls
    • •A common pitfall is conflating verification and validation activities, leading to incomplete testing or a failure to address user needs effectively.
    • •Another mistake is performing V&V too late in the design process, making it costly and difficult to implement necessary changes.
    • •Failure to adequately document V&V activities can result in regulatory non-compliance during audits or submissions.
    • •Not considering cybersecurity V&V throughout the product lifecycle can lead to vulnerabilities in connected MedTech devices.
    • •Over-reliance on software simulations without sufficient hardware testing for embedded systems is a frequent pitfall in software V&V.
    Open full page
    Select term B to compare.

    Why compare MedTech terms side by side?

    MedTech terminology is full of pairs that look interchangeable but carry very different regulatory, clinical, and commercial consequences. Picking the wrong framework, pathway, or standard early in a project can add months to a submission, invalidate clinical evidence, or trigger an audit finding. Side-by-side comparison is the fastest way to surface those differences before they become costly mistakes.

    Each comparison on this page pulls from the same vendor-neutral, sourced definitions used throughout the MedTech Terms glossary. You see the one-line summary, the formal definition, why it matters in practice, common pitfalls, and the primary sources (FDA guidance, EU MDR/IVDR articles, ISO/IEC standards, MDCG documents, IMDRF principles) that back each entry. That makes the comparison defensible in regulatory strategy memos, design reviews, and submission narratives.

    Common comparison patterns

    How to use this tool

    Pick term A and term B from the dropdowns, or click a preset above. The URL updates with both slugs so you can bookmark or share the exact comparison with a colleague, a notified body reviewer, or your regulatory consultant. Click Open full page on either side for the complete entry, including FAQs, related terms, and the full citation list. If you are not sure which term to start with, browse the Categories view or the A-Z index.

    MedTech Terms is a vendor-neutral community resource sponsored by Blue Goat Cyber. Definitions are written for educational use and are not legal or regulatory advice.