    Quality & Risk

    Risk Acceptability Matrix

    In one line
    Pre-defined matrix mapping severity × probability combinations to acceptable, ALARP, or unacceptable risk.
    Definition
    Required by ISO 14971, the risk acceptability matrix is the manufacturer's policy on how risk levels translate into action. It must be defined before risk evaluation and applied consistently across the risk management file. The matrix usually combines severity levels with probability levels and marks regions as acceptable, needing further reduction or unacceptable. ISO 14971:2019 requires the policy for establishing criteria to be set by top management and based on applicable standards, the state of the art and stakeholder concerns. Note that the EU MDR requires risks to be reduced as far as possible, which affects how a matrix's acceptable region is used for EU devices.
    Why it matters
    Notified Bodies frequently challenge inconsistent or unjustified matrices; aligning to ISO/TR 24971 examples is a defensible starting point.
    Common pitfalls
    • •Failing to define the risk acceptability matrix before conducting risk evaluation can lead to biased assessments and non-compliance with ISO 14971.
    • •Inconsistently applying the risk acceptability criteria across different hazards or use scenarios will result in a non-compliant risk management file.
    • •Defining overly aggressive or overly conservative risk acceptability criteria without justification can lead to regulatory scrutiny or an unmarketable device.
    • •Not documenting the rationale for the chosen risk acceptability criteria is a common audit finding.
    • •Using a generic risk matrix without tailoring it to the specific device and its intended use is a significant oversight.
    Open full page
    Select term B to compare.

    Why compare MedTech terms side by side?

    MedTech terminology is full of pairs that look interchangeable but carry very different regulatory, clinical, and commercial consequences. Picking the wrong framework, pathway, or standard early in a project can add months to a submission, invalidate clinical evidence, or trigger an audit finding. Side-by-side comparison is the fastest way to surface those differences before they become costly mistakes.

    Each comparison on this page pulls from the same vendor-neutral, sourced definitions used throughout the MedTech Terms glossary. You see the one-line summary, the formal definition, why it matters in practice, common pitfalls, and the primary sources (FDA guidance, EU MDR/IVDR articles, ISO/IEC standards, MDCG documents, IMDRF principles) that back each entry. That makes the comparison defensible in regulatory strategy memos, design reviews, and submission narratives.

    Common comparison patterns

    How to use this tool

    Pick term A and term B from the dropdowns, or click a preset above. The URL updates with both slugs so you can bookmark or share the exact comparison with a colleague, a notified body reviewer, or your regulatory consultant. Click Open full page on either side for the complete entry, including FAQs, related terms, and the full citation list. If you are not sure which term to start with, browse the Categories view or the A-Z index.

    MedTech Terms is a vendor-neutral community resource sponsored by Blue Goat Cyber. Definitions are written for educational use and are not legal or regulatory advice.