    Manufacturing

    Process Capability and Cpk

    In one line
    Statistical measures (Cp, Cpk, Pp, Ppk) of how well a process output fits within specification limits.
    Definition
    Cpk relates the distance between the process mean and the nearest spec limit to process variation. Cpk ≥ 1.33 is a common minimum target for critical characteristics; ≥ 1.67 for safety-critical attributes. Ppk uses overall variation and is used for initial capability, while Cpk uses within-subgroup variation for ongoing control. Capability studies need enough data, often at least 30 samples, and data that is roughly normal or analyzed with suitable methods. In process validation, capability targets are tied to the risk of the characteristic and are often set in the PQ protocol.
    Why it matters
    PPAP, supplier scorecards, and process validation routinely use Cpk as a lever; 'Cpk theater' (manipulated samples) is a recurring audit concern.
    Common pitfalls
    • •Using Cpk as the sole metric for process control without considering historical data or other statistical process control tools can be misleading.
    • •Manipulating samples to achieve a favorable Cpk without actual process improvement constitutes "Cpk theater" and can lead to non-conformances during audits.
    • •Assuming a good Cpk guarantees product quality without addressing potential sources of variation or measurement system analysis is a common error.
    • •Failing to establish appropriate Cpk targets based on risk and product criticality can result in inadequate process control for critical attributes.
    • •Not periodically re-evaluating Cpk after process changes or over time can lead to a loss of process control without detection.
    Open full page
    Select term B to compare.

    Why compare MedTech terms side by side?

    MedTech terminology is full of pairs that look interchangeable but carry very different regulatory, clinical, and commercial consequences. Picking the wrong framework, pathway, or standard early in a project can add months to a submission, invalidate clinical evidence, or trigger an audit finding. Side-by-side comparison is the fastest way to surface those differences before they become costly mistakes.

    Each comparison on this page pulls from the same vendor-neutral, sourced definitions used throughout the MedTech Terms glossary. You see the one-line summary, the formal definition, why it matters in practice, common pitfalls, and the primary sources (FDA guidance, EU MDR/IVDR articles, ISO/IEC standards, MDCG documents, IMDRF principles) that back each entry. That makes the comparison defensible in regulatory strategy memos, design reviews, and submission narratives.

    Common comparison patterns

    How to use this tool

    Pick term A and term B from the dropdowns, or click a preset above. The URL updates with both slugs so you can bookmark or share the exact comparison with a colleague, a notified body reviewer, or your regulatory consultant. Click Open full page on either side for the complete entry, including FAQs, related terms, and the full citation list. If you are not sure which term to start with, browse the Categories view or the A-Z index.

    MedTech Terms is a vendor-neutral community resource sponsored by Blue Goat Cyber. Definitions are written for educational use and are not legal or regulatory advice.