    Post-Market Performance Follow-Up (IVDR)(PMPF)

    In one line
    Continuous process of proactively collecting and evaluating performance data on an IVD after CE marking, analogous to MDR PMCF.
    Definition
    Annex XIII Part B of the IVDR defines PMPF as a continuous process that updates the performance evaluation and is addressed in the manufacturer's post-market surveillance plan. Its objectives include confirming safety and performance throughout the expected lifetime, identifying previously unknown risks, ensuring the continued acceptability of the benefit-risk ratio, and identifying possible systematic misuse.
    Why it matters
    PMPF is not optional: where a manufacturer justifies that PMPF is not applicable, that justification itself must be documented and defended in the Performance Evaluation Report. Common PMPF activities include real-world performance studies, registry participation, external quality assessment (EQA) programme data, and structured user surveys.
    Common pitfalls
    • •A common mistake is treating PMPF as a one-time activity or an optional requirement; manufacturers must have a continuous process and justify its inapplicability if not pursued.
    • •Manufacturers often fail to adequately integrate PMPF into their overall quality management system, leading to fragmented data collection and evaluation processes.
    • •Another misconception is limiting PMPF activities to only complaint data, overlooking proactive measures like real-world studies or user surveys that provide richer performance insights.
    Open full page
    Select term B to compare.

    Why compare MedTech terms side by side?

    MedTech terminology is full of pairs that look interchangeable but carry very different regulatory, clinical, and commercial consequences. Picking the wrong framework, pathway, or standard early in a project can add months to a submission, invalidate clinical evidence, or trigger an audit finding. Side-by-side comparison is the fastest way to surface those differences before they become costly mistakes.

    Each comparison on this page pulls from the same vendor-neutral, sourced definitions used throughout the MedTech Terms glossary. You see the one-line summary, the formal definition, why it matters in practice, common pitfalls, and the primary sources (FDA guidance, EU MDR/IVDR articles, ISO/IEC standards, MDCG documents, IMDRF principles) that back each entry. That makes the comparison defensible in regulatory strategy memos, design reviews, and submission narratives.

    Common comparison patterns

    How to use this tool

    Pick term A and term B from the dropdowns, or click a preset above. The URL updates with both slugs so you can bookmark or share the exact comparison with a colleague, a notified body reviewer, or your regulatory consultant. Click Open full page on either side for the complete entry, including FAQs, related terms, and the full citation list. If you are not sure which term to start with, browse the Categories view or the A-Z index.

    MedTech Terms is a vendor-neutral community resource sponsored by Blue Goat Cyber. Definitions are written for educational use and are not legal or regulatory advice.