    Regulatory

    Notified Body Scrutiny (IVDR Class D)

    In one line
    Additional expert-panel review of a notified body's assessment for certain novel or high-risk Class D devices before certification.
    Definition
    Article 48(6) requires the notified body assessing a Class D device to notify a designated expert panel of its Performance Evaluation Assessment Report (PEAR) where no CS exist for the device, the device is the first of its kind, or the CS have been recently revised. The expert panel may deliver a scientific view within 60 days that the notified body must duly consider.
    Why it matters
    Scrutiny is a checkpoint intended to raise consistency across notified bodies for the highest-risk IVDs (HIV, HCV, blood grouping, novel pathogen assays). It borrows the mechanism of MDR Article 54 CECP scrutiny for implantable Class III devices.
    Common pitfalls
    • •A common misconception is that scrutiny implies the notified body's assessment is incomplete or incorrect, when in fact it is a standard procedural step for certain high-risk devices.
    • •Manufacturers often assume that if their device is not 'novel' or 'first of its kind,' it will bypass scrutiny, but if no common specifications exist or they were recently revised, scrutiny may still apply.
    • •Some believe that expert panel scrutiny is a final decision-making body, but its role is to provide a scientific view for the notified body to consider, not to approve or reject certification directly.
    Open full page
    Select term B to compare.

    Why compare MedTech terms side by side?

    MedTech terminology is full of pairs that look interchangeable but carry very different regulatory, clinical, and commercial consequences. Picking the wrong framework, pathway, or standard early in a project can add months to a submission, invalidate clinical evidence, or trigger an audit finding. Side-by-side comparison is the fastest way to surface those differences before they become costly mistakes.

    Each comparison on this page pulls from the same vendor-neutral, sourced definitions used throughout the MedTech Terms glossary. You see the one-line summary, the formal definition, why it matters in practice, common pitfalls, and the primary sources (FDA guidance, EU MDR/IVDR articles, ISO/IEC standards, MDCG documents, IMDRF principles) that back each entry. That makes the comparison defensible in regulatory strategy memos, design reviews, and submission narratives.

    Common comparison patterns

    How to use this tool

    Pick term A and term B from the dropdowns, or click a preset above. The URL updates with both slugs so you can bookmark or share the exact comparison with a colleague, a notified body reviewer, or your regulatory consultant. Click Open full page on either side for the complete entry, including FAQs, related terms, and the full citation list. If you are not sure which term to start with, browse the Categories view or the A-Z index.

    MedTech Terms is a vendor-neutral community resource sponsored by Blue Goat Cyber. Definitions are written for educational use and are not legal or regulatory advice.