    Regulatory

    IVDR Risk Classes A–D

    In one line
    EU IVDR's risk-based classification: Class A (lowest) to Class D (highest).
    Definition
    Under EU IVDR (2017/746), IVDs are classified A, B, C, or D based on intended purpose and risk to individuals and public health. Class B and above generally require Notified Body involvement. Classification follows seven rules in IVDR Annex VIII. Class D covers the highest risk tests, such as blood screening for transmissible agents; Class C includes companion diagnostics and many cancer and genetic tests; Class B covers moderate risk devices; Class A covers lower-risk products such as specimen receptacles. Self-testing and near-patient devices often fall into a higher class. All classes except non-sterile Class A need a notified body.
    Why it matters
    IVDR shifted ~80% of IVDs into Notified Body oversight (vs. ~20% under IVDD), creating sustained capacity strain. Common Specifications apply for Class D devices and certain analytes.
    Common pitfalls
    • •Misclassifying an IVD can lead to significant regulatory hurdles, including delays in market access or non-compliance penalties.
    • •Underestimating the evidence required for higher risk classes, particularly for Class C and D devices, is a common pitfall.
    • •Failing to consider all aspects of the intended purpose when determining the risk class can result in an incorrect classification.
    • •Neglecting to update the risk classification if the intended purpose or technology of the IVD changes can lead to non-conformity.
    • •Assuming that a device's classification under the IVDD automatically translates to the same class under the IVDR is incorrect due to the stricter IVDR rules.
    Open full page
    Select term B to compare.

    Why compare MedTech terms side by side?

    MedTech terminology is full of pairs that look interchangeable but carry very different regulatory, clinical, and commercial consequences. Picking the wrong framework, pathway, or standard early in a project can add months to a submission, invalidate clinical evidence, or trigger an audit finding. Side-by-side comparison is the fastest way to surface those differences before they become costly mistakes.

    Each comparison on this page pulls from the same vendor-neutral, sourced definitions used throughout the MedTech Terms glossary. You see the one-line summary, the formal definition, why it matters in practice, common pitfalls, and the primary sources (FDA guidance, EU MDR/IVDR articles, ISO/IEC standards, MDCG documents, IMDRF principles) that back each entry. That makes the comparison defensible in regulatory strategy memos, design reviews, and submission narratives.

    Common comparison patterns

    How to use this tool

    Pick term A and term B from the dropdowns, or click a preset above. The URL updates with both slugs so you can bookmark or share the exact comparison with a colleague, a notified body reviewer, or your regulatory consultant. Click Open full page on either side for the complete entry, including FAQs, related terms, and the full citation list. If you are not sure which term to start with, browse the Categories view or the A-Z index.