    Regulatory

    Class D Batch Verification (IVDR)

    In one line
    Pre-release testing of every manufactured batch of Class D IVDs by an EU Reference Laboratory, where designated.
    Definition
    Article 48(5) and Article 100 require that, for Class D devices where Common Specifications or, in their absence, harmonised standards exist and an EU Reference Laboratory has been designated for the device type, the manufacturer must have each batch tested by the EU RL before release. The EU RL verifies that the batch meets the applicable CS or the manufacturer's declared performance.
    Why it matters
    This is one of the strictest controls in EU device regulation and mirrors requirements historically used for blood-typing and infectious-disease screening. Manufacturers must build EU RL turnaround time into release planning and negotiate service-level expectations early.
    Common pitfalls
    • •Manufacturers may underestimate the time and resources required for EU Reference Laboratory (EU RL) testing, leading to significant delays in product release if not planned effectively.
    • •Assuming that internal quality control testing alone is sufficient for Class D devices, neglecting the mandatory independent verification by an EU RL for designated device types.
    • •Failure to engage with designated EU RLs early in the development and regulatory planning process to understand their specific requirements and turnaround times.
    Open full page
    Select term B to compare.

    Why compare MedTech terms side by side?

    MedTech terminology is full of pairs that look interchangeable but carry very different regulatory, clinical, and commercial consequences. Picking the wrong framework, pathway, or standard early in a project can add months to a submission, invalidate clinical evidence, or trigger an audit finding. Side-by-side comparison is the fastest way to surface those differences before they become costly mistakes.

    Each comparison on this page pulls from the same vendor-neutral, sourced definitions used throughout the MedTech Terms glossary. You see the one-line summary, the formal definition, why it matters in practice, common pitfalls, and the primary sources (FDA guidance, EU MDR/IVDR articles, ISO/IEC standards, MDCG documents, IMDRF principles) that back each entry. That makes the comparison defensible in regulatory strategy memos, design reviews, and submission narratives.

    Common comparison patterns

    How to use this tool

    Pick term A and term B from the dropdowns, or click a preset above. The URL updates with both slugs so you can bookmark or share the exact comparison with a colleague, a notified body reviewer, or your regulatory consultant. Click Open full page on either side for the complete entry, including FAQs, related terms, and the full citation list. If you are not sure which term to start with, browse the Categories view or the A-Z index.

    MedTech Terms is a vendor-neutral community resource sponsored by Blue Goat Cyber. Definitions are written for educational use and are not legal or regulatory advice.