    Quality & Risk

    Failure Mode and Effects Analysis(FMEA)

    In one line
    Bottom-up technique for systematically identifying potential failures and their effects.
    Definition
    FMEA evaluates each component or process step to identify failure modes, their causes, effects, and severity, occurrence, and detectability ratings used to prioritize mitigation. Variants include design FMEA (dFMEA), process FMEA (pFMEA), and use FMEA (uFMEA). Each failure mode is commonly scored for severity, occurrence and detection, sometimes multiplied into a risk priority number (RPN), although many MedTech teams prefer action priority tables because RPNs can hide high severity risks. In medical devices, FMEA supports ISO 14971 risk analysis but does not replace it: FMEA is bottom-up and single-fault, so it can miss hazards that arise from normal use or combined faults.
    Why it matters
    FMEA is a useful input to ISO 14971 risk management but is not, by itself, sufficient - ISO 14971 requires a top-down hazard analysis as well.
    Common pitfalls
    • •FMEA alone is not sufficient to meet the requirements of a comprehensive risk management system like that described in ISO 14971, which also requires a top-down hazard analysis.
    • •Focusing solely on FMEA without considering the broader context of system-level hazards can lead to an incomplete risk profile.
    • •Incorrectly assigning severity, occurrence, or detectability ratings can result in an inaccurate prioritization of risks.
    • •FMEA can be perceived as a one-time exercise rather than an iterative process that evolves throughout the device lifecycle.
    • •Failing to link FMEA outputs directly to risk control measures and verification activities can diminish its effectiveness.
    Open full page
    Select term B to compare.

    Why compare MedTech terms side by side?

    MedTech terminology is full of pairs that look interchangeable but carry very different regulatory, clinical, and commercial consequences. Picking the wrong framework, pathway, or standard early in a project can add months to a submission, invalidate clinical evidence, or trigger an audit finding. Side-by-side comparison is the fastest way to surface those differences before they become costly mistakes.

    Each comparison on this page pulls from the same vendor-neutral, sourced definitions used throughout the MedTech Terms glossary. You see the one-line summary, the formal definition, why it matters in practice, common pitfalls, and the primary sources (FDA guidance, EU MDR/IVDR articles, ISO/IEC standards, MDCG documents, IMDRF principles) that back each entry. That makes the comparison defensible in regulatory strategy memos, design reviews, and submission narratives.

    Common comparison patterns

    How to use this tool

    Pick term A and term B from the dropdowns, or click a preset above. The URL updates with both slugs so you can bookmark or share the exact comparison with a colleague, a notified body reviewer, or your regulatory consultant. Click Open full page on either side for the complete entry, including FAQs, related terms, and the full citation list. If you are not sure which term to start with, browse the Categories view or the A-Z index.

    MedTech Terms is a vendor-neutral community resource sponsored by Blue Goat Cyber. Definitions are written for educational use and are not legal or regulatory advice.