    Regulatory

    Clinical Evaluation Consultation Procedure(CECP)

    In one line
    MDR Article 54 procedure in which an independent expert panel reviews the notified body's clinical evaluation assessment for certain Class III and implantable Class IIb devices.
    Definition
    The Clinical Evaluation Consultation Procedure is set out in MDR Article 54 and detailed in Annex IX Section 5.1. It requires the notified body to notify the European Commission of its clinical evaluation assessment report for implantable Class III devices and certain implantable and Class III active devices that administer or remove a medicinal product, so that an independent expert panel can scrutinise the manufacturer's clinical evaluation. The expert panel may issue a scientific opinion within 60 days (extendable), and while the notified body is not bound by it, Annex IX requires the notified body to duly consider the opinion and document its final position in the conformity assessment.
    Why it matters
    CECP does not apply to every Class III or implantable device: Article 54(2) and Annex IX Section 5.1 carve out exceptions for renewals of existing certificates without significant changes, devices already CE-marked and modified only in specific limited ways, and certain well-established device types (as clarified in MDCG guidance). CECP was designed as a check on notified body inconsistency for the highest-risk devices, echoing lessons from cases such as the PIP breast implant scandal that preceded the MDR.
    Common pitfalls
    • •Assuming all Class III implantables automatically require CECP; several exemptions in Article 54(2) and MDCG 2019-3 can remove the obligation.
    • •Underestimating the added 60-plus day timeline for expert panel review when planning a certification schedule for a novel implantable device.
    • •Treating a favorable expert panel opinion as a guarantee of certification; the notified body retains final decision authority and may still require additional clinical evidence.
    Open full page
    Select term B to compare.

    Why compare MedTech terms side by side?

    MedTech terminology is full of pairs that look interchangeable but carry very different regulatory, clinical, and commercial consequences. Picking the wrong framework, pathway, or standard early in a project can add months to a submission, invalidate clinical evidence, or trigger an audit finding. Side-by-side comparison is the fastest way to surface those differences before they become costly mistakes.

    Each comparison on this page pulls from the same vendor-neutral, sourced definitions used throughout the MedTech Terms glossary. You see the one-line summary, the formal definition, why it matters in practice, common pitfalls, and the primary sources (FDA guidance, EU MDR/IVDR articles, ISO/IEC standards, MDCG documents, IMDRF principles) that back each entry. That makes the comparison defensible in regulatory strategy memos, design reviews, and submission narratives.

    Common comparison patterns

    How to use this tool

    Pick term A and term B from the dropdowns, or click a preset above. The URL updates with both slugs so you can bookmark or share the exact comparison with a colleague, a notified body reviewer, or your regulatory consultant. Click Open full page on either side for the complete entry, including FAQs, related terms, and the full citation list. If you are not sure which term to start with, browse the Categories view or the A-Z index.

    MedTech Terms is a vendor-neutral community resource sponsored by Blue Goat Cyber. Definitions are written for educational use and are not legal or regulatory advice.