Crosswalks

    US vs EU medical device cybersecurity, side by side

    Since March 2023 the US has a device-specific cybersecurity law, section 524B of the FD&C Act, enforced through FDA's premarket review. The EU has no separate device cybersecurity law; cybersecurity sits inside the MDR's general safety and performance requirements and is explained by MDCG guidance. Manufacturers selling in both markets can usually build one security file, but the US expects it to be submitted and the EU expects it to be audited.

    By Christian Espinosa, Founder, Blue Goat Cyber

    Legal basis and scope

    TopicUnited States (FDA)European Union (MDR)What differs
    Primary lawFD&C Act section 524B, for cyber devices21 U.S.C. 360n-2MDR GSPRs on IT security and protection against unauthorised accessMDR Annex I, sections 14.2(d), 17.2, 17.4, 18.8524B applies to devices that include software, can connect to the internet and could be vulnerable. The MDR requirements apply to all devices with software or electronic programmable systems.
    Main guidanceFDA premarket cybersecurity guidance (Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions)FDA guidanceMDCG 2019-16, Guidance on Cybersecurity for medical devicesMDCG 2019-16Both are guidance, not law, but reviewers and notified bodies use them as the yardstick.
    Related horizontal lawsNone specific to devices beyond 524B-Cyber Resilience Act excludes MDR devices; NIS2 can apply to manufacturers as entitiesRegulation (EU) 2024/2847, Directive (EU) 2022/2555Check NIS2 national transposition for whether your company is in scope.

    Premarket evidence

    TopicUnited States (FDA)European Union (MDR)What differs
    Secure development processSecure product development framework (SPDF)524B(b)(2)Secure lifecycle per IEC 81001-5-1MDR Annex I 17.2, harmonisation pendingIEC 81001-5-1 satisfies both expectations in practice and is FDA-recognized.
    Threat and risk analysisThreat modeling and security risk managementFDA premarket guidanceSecurity risk management integrated with ISO 14971MDCG 2019-16 section 3AAMI TIR57 is the common method for linking security risk to safety risk.
    Software bill of materialsSBOM required in the submission524B(b)(3)Not named in the MDR; expected by notified bodies as good practiceMDCG 2019-16An SBOM built for FDA can be reused in the EU technical documentation.
    Security testingPenetration testing, fuzzing and vulnerability scanning evidenceFDA premarket guidanceVerification and validation of security controlsMDR Annex I 17.2, MDCG 2019-16Report formats differ little; the US reviewer reads the results, the EU notified body audits them.
    LabelingCybersecurity information for users and an end-of-support dateFDA premarket guidanceMinimum IT requirements and security information in the IFUMDR Annex I 23.4(ab)Both expect users to be told how to operate the device securely.

    Post-market

    TopicUnited States (FDA)European Union (MDR)What differs
    Vulnerability handlingPlan to monitor, identify and address vulnerabilities, including coordinated disclosure524B(b)(1)Post-market surveillance must cover security; vulnerabilities handled through PMS and vigilanceMDR Articles 83 to 92, MDCG 2019-16Both expect a public way for researchers to report vulnerabilities.
    Updates and patchesUpdates and patches on a reasonably justified cycle, and out of cycle for critical risks524B(b)(2)Changes assessed under the QMS and, if significant, by the notified bodyMDR Article 120 and Annex IXIn both markets, a security patch rarely needs a new submission, but document the assessment.
    ScoringCVSS commonly used, with a medical device rubricFDA-recognized CVSS rubric (MITRE)CVSS commonly used; no mandated methodMDCG 2019-16Translate exploitability into patient harm before prioritising fixes.
    Incident reportingReported through MDR or corrections and removals when safety is affected21 CFR Parts 803 and 806Serious incident or FSCA under vigilance when safety is affectedMDR Articles 87 and 89A vulnerability alone is not always reportable; the safety impact decides.

    Primary sources

    Educational summary, not legal advice. Always check the current text of the law and guidance.